Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Product picture of: Oxymax W COS71

Oxymax W COS71

COS71
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task
  • Application

    Boiler feed water, power plant,

  • Characteristic

    Amperometric 3-electrode sensor for the trace measuring range.

  • Measurement range

    0,001 - 20mg/l

  • Measuring principle

    Closed (membrane covered) amperometric 3-electrode system)

  • Design

    - Automatic adjustment for digitale or analog communication depending on the transmitter.- Storage of the calibration datas in the sensor.- Extreme stable membrane.

  • Material

    Sensor body : Stainless steel 1.4571Membrane cap : POM

  • Dimension

    Diameter : 40mm
    (1.56inch)Length : 220mm
    (8.58inch)

  • Process temperature

    max. 50°C
    (max.122°F)

  • Process pressure

    max. 10bar
    (max.145psi)

  • Temperature sensor

    Integrated NTC-sensor

  • Connection

    Process connection : G1"Cable : fixed cable or TOP68.